ZIMMER BIOMET, INC. PERSONA CPS TIBIAL ARTICULAR SURFACE PROVISIONAL PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Patient Involvement (2645)
Event Date 08/17/2017
Event Type  malfunction  
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed. Visual inspection of the returned cps tibial articular surface provisional exhibits signs of repeated use and the anterior section of the part has been ripped through. Provisionals are subjected to wear and damage due to use over time and this is addressed in the package insert. Device history record (dhr) was reviewed and no discrepancies were found. Root cause of the event could not be determined. If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly. Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
(b)(6). The complaint device has been returned, but the device investigation has not yet been completed. Once the evaluation is completed, a supplemental medwatch 3500a will be submitted. If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly. Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the provisional was fratured and returned. This incident did not occur during a surgery and no adverse events have been reported as a result of this malfunction as there is no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePERSONA CPS TIBIAL ARTICULAR SURFACE PROVISIONAL
Type of DevicePROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7044707
MDR Text Key256154021
Report Number0001822565-2017-07806
Device Sequence Number1
Product Code OIY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK123459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 04/09/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberN/A
Device Catalogue Number42527600410
Device Lot Number63693450
Other Device ID Number(01) 00889024245457
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/24/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown
Removal/Correction NumberN/A

-
-