Model Number 174301
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/20/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Visual inspection of the device revealed signs of clinical use.Debris is present on the blades of the device.The device was inspected under magnification and the blades revealed wear.Functional inspection revealed the blades to be dull and unable to provide a clean cut.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root cause(s) are: user attempts to cut staples, clips, or non-soft tissue; wear to blades due to clinical use.The instructions for use (ifu) state: do not directly apply current to staples or clips.Instruments were designed for cutting soft tissue.Attempting to cut staples or clips may damage the instrument.If cutting of staples or clips is attempted, damage to instrument may occur.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported the laparoscopic scissors were dull.The device was replaced.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
MDR Report Key7044911
Report Number0001056128-2017-00138
Device Sequence Number16181977
Product Code NUJ
UDI-Device Identifier00885825012564
UDI-Public(01)00885825012564(17)190721(10)4746828
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K012700
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/21/2019
Device Model Number174301
Device Catalogue Number174301RR
Device Lot Number4746828
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2017
Initial Report FDA Received Date11/20/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-