• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVE DIVISION CG FUTURE BAND; RING, ANNULOPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVE DIVISION CG FUTURE BAND; RING, ANNULOPLASTY Back to Search Results
Model Number 638BL
Medical Device Problem Code Perivalvular Leak (1457)
Health Effect - Clinical Code Mitral Regurgitation (1964)
Date of Event 10/19/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that this 28 mm mitral bioprosthetic band was implanted and explanted during the same procedure.Post implant of the 28 mm, after tied and sutured into place, the physician noted that the leaflets coapted well and there was essentially no regurgitation.A post implanted transesophageal echocardiogram (tee) was done which showed visually the valve looked completely normal, once a doppler was placed on the valve, it was evident that although improved, there was significant mitral regurgitation and the anterior leaflet appeared to be leaking, but the posterior leaflet was not prolapsing.The physician decided that there was too much regurgitation to leave the band in place, so the patient was placed back on bypass.The 28 mm band was removed and successfully replaced with a 29 mm bioprosthetic valve.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CG FUTURE BAND
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC HEART VALVE DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key7045113
Report Number2025587-2017-02195
Device Sequence Number16172595
Product Code KRH
UDI-Device Identifier00613994758774
UDI-Public00613994758774
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K052860
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/22/2021
Device Model Number638BL
Device Catalogue Number638BL28
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2017
Initial Report FDA Received Date11/20/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
-
-