| Model Number |
EN28142400 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Codes |
No Information (3190); No Code Available (3191)
|
| Date of Event |
11/07/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information of conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
According tot he available information, 14 cm titan touch was explanted because it was too short.Patient was revised with 20 cm titan touch.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up mdr is created to document additional information.
|
| |
|
Event or Problem Description
|
|
Additional information received: difficulty to pump.The device is not available for return to coloplast.Device was replaced with another titan.No infection.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up mdr is created to document the conclusion of the investigation.A device was received in a condition making evaluation impossible.However, because quality's examination may not conclusively confirm or disprove the report of incorrect sizing, quality accepts the physician's observations of such as the reason for surgical intervention.Quality reviewed the manufacturing records for this device and confirmed that there were no discrepancies that would have contributed to this complaint and verified that the devices from this lot met all specifications prior to release.Review of nonconforming reports revealed no nonconformance with this lot that would have contributed to the event.A review of the complaint database revealed no significant trends in complaints of this type for lot 4028728.The most likely root cause of the event cannot be determined at this time.If additional information is received, quality will re-evaluate this complaint in accordance to procedures.Management routinely reviews events such as this and monitors complaint levels.Additionally, events of this type are captured in the product risk documentation.Based on this information, no further corrective action is required at this time.
|
| |
|