| Model Number |
1435- |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/08/2015
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One cdc adapter was not returned for evaluation.Review of the manufacturing documentation confirmed that the associated device met all requirements for release.The reported event could not be confirmed since log files were not available for analysis.Analysis of the device could not be performed since the device was not returned for evaluation.Analysis could not be performed as the device was not returned.This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.This event was assessed and is being reported as part of a retrospective review of events, which was in response to an update to the mdr decision criteria.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient noticed damage to dc adapter connector.The damage to the connector prevented a secure connection.The patient discontinued use of the adapter.There were no reported patient consequences associated with the event.No further information was provided.
|
| |
|