| Catalog Number |
VA80104R |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the subclavian vein, the pta balloon catheter allegedly could not be retracted through the sheath.The balloon and sheath were removed as a single unit.The procedure was concluded, and no further treatment was provided.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Conclusion: the device was returned within a 6fr introducer sheath.A visual inspection found the balloon material to be bunching at the distal tip of the introducer sheath.The distal tip of the sheath was examined under microscopic magnification (12.5x) and the tip was found to be damaged and flared.Therefore, the investigation is confirmed for the reported sheath-related retraction issues.The definitive root cause for the identified retraction problems could not be determined based upon the available information.It is unknown if patient and/or procedural issues contributed to the reported event.Labeling review: the current ifu (instructions for use) states: warnings: when the catheter is exposed to the vascular system, it should be manipulated while under high-quality fluoroscopic observation.Do not advance or retract the catheter unless the balloon is fully deflated.If resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.Do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.Precautions: if resistance is felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and guidewire/introducer sheath as a single unit.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the subclavian vein, the pta balloon catheter allegedly could not be retracted through the sheath.The balloon and sheath were removed as a single unit.The procedure was concluded, and no further treatment was provided.There was no reported patient injury.
|
| |
|