| Catalog Number |
DP-SDP001 |
| Medical Device Problem Codes |
Defective Component (2292); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
10/01/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This event is currently under investigation.A follow-up report will be generated upon the completion of the investigation.
|
| |
|
Event or Problem Description
|
|
As reported to customer relations: "the doppler was initially working when it was put into the patient, but the crystal stopped working after the surgery was complete.The patient had to go back for a second surgery in order to replace the doppler.This required the patient to undergo a second general anesthesia, increased the risk of infection, increased the risk of neurovascular issues and delayed discharge from hospital.".
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown or unavailable.(b)(4).Additional mfg narrative - notes - added summary of investigation investigation/evaluation summary: the returned device was inspected under magnification and the customer's complaint could be confirmed.A functional test however could not be performed as the probe wire was cut away on the returned unit.No other nonconformities or anomalies were observed on the device.A review of the device history record and manufacturing and quality control records was conducted.There is no evidence that defective product was manufactured using current drawings and device specifications.Complaints will continue to be monitored according to established complaint monitoring procedures.
|
| |
|