• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK VASCULAR INC COOK-SWARTZ DOPPLER PROBE; ITX TRANSDUCER, ULTRASONIC, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK VASCULAR INC COOK-SWARTZ DOPPLER PROBE; ITX TRANSDUCER, ULTRASONIC, DIAGNOSTIC Back to Search Results
Catalog Number DP-SDP001
Medical Device Problem Codes Defective Component (2292); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/01/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).This event is currently under investigation.A follow-up report will be generated upon the completion of the investigation.
 
Event or Problem Description
As reported to customer relations: "the doppler was initially working when it was put into the patient, but the crystal stopped working after the surgery was complete.The patient had to go back for a second surgery in order to replace the doppler.This required the patient to undergo a second general anesthesia, increased the risk of infection, increased the risk of neurovascular issues and delayed discharge from hospital.".
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.(b)(4).Additional mfg narrative - notes - added summary of investigation investigation/evaluation summary: the returned device was inspected under magnification and the customer's complaint could be confirmed.A functional test however could not be performed as the probe wire was cut away on the returned unit.No other nonconformities or anomalies were observed on the device.A review of the device history record and manufacturing and quality control records was conducted.There is no evidence that defective product was manufactured using current drawings and device specifications.Complaints will continue to be monitored according to established complaint monitoring procedures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK-SWARTZ DOPPLER PROBE
Common Device Name
ITX TRANSDUCER, ULTRASONIC, DIAGNOSTIC
Manufacturer (Section D)
COOK VASCULAR INC
1186 montgomery lane
vandergrift PA 15690
MDR Report Key7048212
Report Number2522007-2017-00043
Device Sequence Number1450574
Product Code ITX
UDI-Device Identifier00827002213630
UDI-Public(01)00827002213630(17)200731(10)N148395
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K022649
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberDP-SDP001
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 10/26/2017
Supplement Date Received by Manufacturer10/26/2017
Initial Report FDA Received Date11/21/2017
Supplement Report FDA Received Date02/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-