| Catalog Number |
DP-SDP001 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
10/01/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.
|
| |
|
Event or Problem Description
|
|
As reported to customer relations: "the doppler was initially working when put in the patient, but crystal detached/slipped after the surgery was complete.The patient had to go back for a second surgery in order to replace the doppler.This required the patient to undergo a second general anesthesia, increased the risk of infection, increased the risk of neurovascular issues and delayed discharge from hospital.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation/evaluation summary: the crystal/cuff was not returned with the doppler probe, only a section of the probe wire.The returned piece was inspected under magnification.A functional test could not be performed since a complete device was not returned.No other nonconformities or anomalies were observed on the device.A review of the device history record and manufacturing and quality control records was conducted.There is no evidence that defective product was manufactured using current drawings and device specifications.Complaints will continue to be monitored according to established complaint monitoring procedures.
|
| |
|