| Model Number |
2110-0039 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
| Date of Event |
10/30/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; locking screw broke in plate.Update 10/31/2017: event occured during surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Medical products: catalog# 48804316a, lot# m-61282.Method: visual inspection;device history review; complaint history review; risk assessment; results: manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.Conclusion: the probable root cause is lack of preparation of the screw hole as recommended by the stg.
|
| |
|
Event or Problem Description
|
|
It was reported that; locking screw broke in plate.Update 10/31/2017: event occured during surgery.
|
| |
|