• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. M2A-MAGNUM 42-50 TPR INSRT STD; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. M2A-MAGNUM 42-50 TPR INSRT STD; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Difficult to Remove (1528)
Patient Problem No Code Available (3191)
Event Date 10/26/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Additional concomitant medical products: (b)(4), bi-metric/x por nc lat 14x150, (b)(4); (b)(4), m2a-magnum pf cup 56odx50id, (b)(4); (b)(4), m2a-magnum mod hd sz 50mm, (b)(4).Customer has been indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2017.Hospital discarded as biohazard.
 
Event Description
It was reported that the taper and head were cold welded to the stem, which resulted in a delay of removal for 30-45 minutes.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.Device history record was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no trends were identified.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M2A-MAGNUM 42-50 TPR INSRT STD
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7050473
MDR Text Key92684332
Report Number0001825034-2017-10206
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK042037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date03/31/2015
Device Model NumberN/A
Device Catalogue Number139256
Device Lot Number729100
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2017
Initial Date FDA Received11/21/2017
Supplement Dates Manufacturer Received12/12/2017
Supplement Dates FDA Received12/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/04/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age63 YR
-
-