Model Number MMT-1715K |
Device Problems
No Display/Image (1183); Improper or Incorrect Procedure or Method (2017)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/09/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
The customer reported via phone call that their insulin pump had a blank display.The customer¿s blood glucose level was 150 mg/dl.The customer states changed the battery and pump hit the table and hit the floor.The customer also states their pump beeps and the screen is lit up, but it won't do anything.Troubleshooting was performed for blank display and noticed display is flashing.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
Device received with blank flashing white lcd due to cracked on lcd controller.Unable to verify missing segments or partial display due to blank flashing white lcd.Unable to performed displacement, rewind, seating and basic occlusion due to blank display.
|
|