• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALLINCKRODT MEDICAL SHILEY; TUBE, BRONCHIAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT MEDICAL SHILEY; TUBE, BRONCHIAL (W/WO CONNECTOR) Back to Search Results
Model Number 125035
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/02/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a communication regarding a customer that could take out the air from the balloon.The customer indicated that there was no patient involvement associated to this event.
 
Additional Manufacturer Narrative
The sample was returned and visual examination showed that the shape returned unit had been altered using the stylet wire.Visual examination shows that there is a clear gel like material in the main stem of the tubing.The stylet wire was removed and inflation/deflation testing was successful.There was no fault found with the device.The probable root cause is that the cuff test was carried out when the shape of the tube was incorrectly altered with the stylet wire.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE, BRONCHIAL (W/WO CONNECTOR)
Manufacturer (Section D)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
MDR Report Key7053518
Report Number8020889-2017-05189
Device Sequence Number1211515
Product Code BTS
UDI-Device Identifier30884522008734
UDI-Public30884522008734
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K771219
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2019
Device Model Number125035
Device Catalogue Number125035
Device Lot Number201411522X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2018
Initial Date Received by Manufacturer 11/04/2017
Supplement Date Received by Manufacturer01/24/2018
Initial Report FDA Received Date11/22/2017
Supplement Report FDA Received Date02/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-