| Model Number |
125035 |
| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that prior to use, during inspection, the cuff of an endobronchial tube could not be deflated completely.The customer reported that there was no patient involvement.The sample is expected to be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary one sample was received for analysis and the reported issue could not be confirmed.A visual inspection was performed and it was observed all the components were assembled according to manufacturing process.A inflation/deflation test was performed, air was applied to the cuff and it was observed the cuff held the air and no deformations or anomalies were observed.If information is provided in the future, a supplemental report will be issued.
|
| |
|