Model Number 1435-
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/01/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product event summary: one controller dc (cdc) adapter was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.The reported event was confirmed via functional testing.Analysis of the device revealed that the device failed to meet specifications; the device passed visual examination but failed functional testing due to a break in the cable connection in the dc lighter connector.The confirmed malfunction is related to the reported event.The most likely root cause of the cdc adapter's failure can be attributed to wear at the strain relief, as a result of excessive bending, likely contributing to fraying or breaking of a wire.This event was assessed and is being reported as part of a retrospective review of events, which was in response to an update to the mdr decision criteria.This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that controller dc (cdc) adapter no longer worked, and different outlets were tried without luck.The cdc adapter was exchanged.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM- CONTROLLER DC ADAPTER
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key7054327
Report Number3007042319-2017-04222
Device Sequence Number1210857
Product Code DSQ
UDI-Device Identifier00888707000109
UDI-Public00888707000109
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 11/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1435-
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2017
Initial Report FDA Received Date11/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-