Model Number 72201919S
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/17/2017
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was overheating.No patient injury was reported.
 
Additional Manufacturer Narrative
The device was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of overheating could not be reproduced.Product passed functional testing during 6 hour burn-in with no overheating or signal loss.Video image remained constant throughout burn-in.All functions perform as expected.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVICE, CCU, HIGH DEF, 560P, 2NDGEN
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key7054887
Report Number1643264-2017-01767
Device Sequence Number1466320
Product Code GCJ
UDI-Device Identifier03596010645517
UDI-Public(01)03596010645517
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K070266
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 01/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72201919S
Device Catalogue Number72201919S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2017
Initial Date Received by Manufacturer 11/17/2017
Supplement Date Received by Manufacturer01/11/2018
Initial Report FDA Received Date11/22/2017
Supplement Report FDA Received Date01/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-