• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA MECTALIF POSTERIOR INTERBODY FUSION DEVICE PEEK/TI 11X22X8 L5°

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA MECTALIF POSTERIOR INTERBODY FUSION DEVICE PEEK/TI 11X22X8 L5° Back to Search Results
Catalog Number 03.27.033
Medical Device Problem Code Device Issue (2379)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/26/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
On 07 november 2017 the r and d project manager performed a preliminary investigation and commented as follows: no pictures of the devices or of the devices assembled with the related inserter have been provided.For this reason based on the description mentioned in the complaint, it seem that there are problem during the implant assembling.Anyway it's not possible to identify a route cause of the event.The inserter was use to finalize the surgery with a different implant lot, for this reason the functionality of the inserter seems to be proven.On the other side the cage was produce according to the specification and no annotation have been filed in the route card.For all the reasons mentioned above it's not possible to identify a specific route cause of the event.Batch review performed on 20 november 2017.(b)(4).
 
Event or Problem Description
Two posterior cages were not able to fix straight to the handle (lot unknown).The surgeon replaced with new cages of a different lot.The surgery was prolonged about 10 minutes.The issue consisted in a suboptimal coupling between cages and instruments (no implant/ instrument deformation nor damage).The surgery was completed with the same instruments.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MECTALIF POSTERIOR INTERBODY FUSION DEVICE PEEK/TI 11X22X8 L5°
Common Device Name
POSTERIOR INTERBODY FUSION DEVICE
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
MDR Report Key7057542
Report Number3005180920-2017-00702
Device Sequence Number1157106
Product Code MAX
UDI-Device Identifier07630030831751
UDI-Public07630030831751
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133192
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/19/2021
Device Catalogue Number03.27.033
Device Lot Number1620432
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/26/2017
Initial Report FDA Received Date11/24/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-