| Model Number |
FG-06553-001A |
| Medical Device Problem Codes |
Entrapment of Device (1212); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/03/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The root cause of the incident is unknown at the time of this report.The complaint device was not returned for evaluation yet, therefore the investigation and root cause analysis could not be performed.The dhr of the affected lot 632803 and related sub-assemblies lots 636794 and 635707 were reviewed without any non-conformities found.The devices passed all 100% in process and aql based qc inspections.The manufacturer informed manufacturing operators and qc inspectors about the complaint.The event has not caused any harm to the patient.No remedial actions were initiated at the time of this report.
|
| |
|
Event or Problem Description
|
|
The following incident description was reported by the customer: "a fortress 6f introducer sheet was performing normally during a pta crossover intervention.During removal of the sheet the interventional physician (b)(6) had a hard time to complete the procedure.The sheet elongated itself and was finally removed in one piece.There was no complication to the patient or any parts remained in the body.".
|
| |
|
Additional Manufacturer Narrative
|
|
Sample evaluation: the sample was returned for investigation.The reported elongation of the product was confirmed.The tip of the sheath was not damaged.The elongation of the sheath begins at approximately 2cm from the distal end, which is probably the spot which caused the friction.Od of the sheath was measured with ring gage used for inspection of 6fr (2,88mm) - the sheath passed through the ring gauge without any complications - it is confirmed, that the product was within the specification.According to the information provided by the customer, there were no difficulties during preparation and insertion of the sheath into the patient the customer informed cmi, that the dilator was not used during removal which is inconsistent with the instructions in ifu.The end user did not follow the instructions in the ifu - the dilator was not used during the removal, which is the most likely root cause.Other potential root causes could be e.G.Patient anatomy, failure of other devices used during the procedure etc.
|
| |
|
Search Alerts/Recalls
|