Model Number A60P9B IS3000 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/31/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Isi has not received the unit involved with this complaint, therefore, the root cause of the customer reported failure mode cannot be determined.A follow-up mdr will be submitted if additional information is received.This complaint is being reported due to a da vinci system malfunction rendering the da vinci system unavailable for use after the start of a surgical procedure.Although no patient harm occurred, if this malfunction were to recur it could cause or contribute to an adverse event.
|
|
Event Description
|
It was reported that during a da vinci-assisted myomectomy procedure, the second surgeon side console (ssc) froze.The customer was informed by the intuitive surgical, inc.(isi) technical support engineer (tse) that it could be due to an error 75.The customer still had one hour remaining for the case.The procedure was completed using the first ssc.There was no report of patient harm, adverse outcome or injury.An isi field service engineer (fse) was dispatched to the facility and was able to reproduce the problem.The fse replaced the touch screen to resolve the issue.
|
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) has received the touch screen computer involved with this complaint and completed the device evaluation.Failure analysis investigation confirmed the reported failure.Visual inspection was performed and a flat cable was found touching the sheet metal.The part was installed to pca system and at startup, the touchpad was blank and an error 75 popped up.
|
|
Search Alerts/Recalls
|