MEDTRONIC NEUROSURGERY DELTA SHUNT KIT, REGULAR, PERFORMANCE LEVEL 1.5; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 25132-5
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/17/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during pre-operative testing, it was determined that the delta chamber was leaking.The product was replaced and there was no injury to the patient.
 
Additional Manufacturer Narrative
The returned valve was patent.The valve met the requirements for siphon, pressure-flow, and pre-implantation testing.The valve did not meet the requirements for reflux testing.There was proteinaceous debris observed within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open resulting in fluid reflux.Instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ the returned valve also did not meet the requirements for leak due to a tear in the top of the delta chamber.It is unknown how or when this damage occurred.The ifu caution that ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting, or crushing of components.¿ the ifu also caution that ¿care should be taken in handling the valves as silicone has a low cut and tear resistance.All valves are 100% tested at the time of manufacture.Approximately 23 cm of the ventricular catheter was returned.It met the requirements for patency and leak testing.Approximately 90 cm of the peritoneal catheter was returned.It met the requirements for patency and leak testing.All catheters are 100% inspected at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA SHUNT KIT, REGULAR, PERFORMANCE LEVEL 1.5
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7060984
Report Number2021898-2017-00595
Device Sequence Number16623364
Product Code JXG
UDI-Device Identifier00643169466418
UDI-Public00643169466418
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K902783
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Model Number25132-5
Device Catalogue Number25132-5
Device Lot NumberE02695
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2017
Supplement Date Received by Manufacturer04/19/2018
Initial Report FDA Received Date11/27/2017
Supplement Report FDA Received Date04/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age33 YR
-
-