| Catalog Number |
70468580 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that patient was pre-operatively diagnosed with scoliosis and underwent an unspecified surgery.Post-op, screw was found broken at s2.The screw remained in the patient.Revision surgery was not scheduled as there were no issues noted.
|
| |
|
Additional Manufacturer Narrative
|
|
X-ray review: post-op x-rays for t2- iliac fixation provided.Hardware placement appears appropriate and proper contouring of rods and correction of sagittal balance appears to have happened.No pre-op films available but as per these images, there are fractures of both iliac screws.From images, we are unable to assess fusion status.If information is provided in the future, a supplemental report will be issued.
|
| |
|