• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL,INC. DA VINCI SI SURGICAL SYSTEM; ENDOSCOPIC INSTRUMENT CONTROL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL,INC. DA VINCI SI SURGICAL SYSTEM; ENDOSCOPIC INSTRUMENT CONTROL SYSTEM Back to Search Results
Model Number IS3000 A60P9B
Device Problems Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/02/2017
Event Type  malfunction  
Manufacturer Narrative
Intuitive surgical, inc.(isi) received the unit involved with this complaint and completed the device evaluation.Failure analysis reproduced the customer reported failure.The illuminator was installed into the system and it failed to power on with errors.This complaint is being reported due to a da vinci system malfunction rendering the da vinci system unavailable for use after the start of a surgical procedure.Although no patient harm occurred, if this malfunction were to recur it could cause or contribute to an adverse event.
 
Event Description
It was reported that prior to the start of a da vinci-assisted surgical procedure, an error occurred at startup.The intuitive surgical, inc.(isi) technical support engineer (tse) guided the isi clinical sales representative (csr) through troubleshooting, however, errors persisted.It was recommended to use an external illuminator to continue with the setup.Isi followed up with the reporter and obtained additional information about the complaint.The issue occurred when the patient was being put under anesthesia.The csr believes that additional anesthesia was given due to the time that it took to find an external illuminator.There were no reports of intraoperative or post-operative complications.An isi field service engineer (fse) was dispatched to the facility and was able to reproduce the reported failure and the error was confirmed through the system logs.To resolve the issue the fse replaced the illuminator.The illuminator is a component of the da vinci system that contains a high intensity light source to illuminate the surgical site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DA VINCI SI SURGICAL SYSTEM
Type of Device
ENDOSCOPIC INSTRUMENT CONTROL SYSTEM
Manufacturer (Section D)
INTUITIVE SURGICAL,INC.
sunnyvale CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC.,
950 kifer rd
sunnyvale CA
Manufacturer Contact
tabitha reed
950 kifer rd
sunnyvale, CA 
4085232420
MDR Report Key7065775
MDR Text Key93724142
Report Number2955842-2017-00802
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K081137
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIS3000 A60P9B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received11/09/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-