The subject clv-s40pro was returned to omsc.However, omsc could not conduct the verification to reproduce the phenomenon with non-olympus lamp mounted in the subject clv-s40pro, because non-olympus lamp was not returned.Omsc mounted olympus lamp in the subject clv-s40pro, and confirmed the operation.Then the subject clv-s40pro was operated normally.As a result of the investigation, omsc concluded that there was no abnormality with the subject clv-s40pro and that this event was caused by using the non-olympus lamp.The clv-s40pro instruction manual states that never install a lamp that has not been approved by olympus.There were no further details provided.If significant additional information is received, this report will be supplemented.
|
The followings were reported to olympus medical systems corp.(omsc).-after 30 minutes from starting an abdominal laparoscopic salpingo resection, the lamp in the subject clv-s40pro was broken.-the user replaced the endoscopic system including the subject clv-s40pro with the spare system, and completed the procedure.-there was no report of the patient¿s injure regarding this event.-the broken lamp was not olympus manufactured.
|