| Model Number |
VLOCL0014 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
To date, the incident sample has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained, a follow-up report will be submitted.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the loop broke when the first stitch got through and the thread was pulled up with a slight force.The procedure was completed with another device.The status of the patient is no problem.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the sample noted one suture and one needle.The product was received with the loop open.It was observed, under magnification, that the loop appeared to have split under tension.Upon microscopic inspection, evidence of material transfer was found at the weld site of the loop.As no sealed products were returned, functional testing was precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.As no sealed products were returned, functional testing was precluded.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|