| Catalog Number |
45-80100 |
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
11/01/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "k-wire was driven in to bone with stryker cd8 cordless driver and wire collet (facility owned).The k-wire broke off in the fibula.Case was an orif right fibula.The k-wire was barely protruding posterior and medial in the fibula.The capital fragment of the wire was removed but broken tip was stuck in the bone and unable to be removed easily.Tip protruding was too short to grip with the wire collet and further complicated by its position being difficult to access.Surgeon attempted to cut the tip to no avail.Surgeon attempted to bend the wire up and into the bone to prevent soft tissue irritation to no avail.Wire was ultimately removed by making multiple holes in the bone with a k-wire adjacent to the wire to loosen and free it to the point where it could be gripped and pulled free." delay in surgery and multiple holes made in the bone with another k-wire.Surgical delay of 30-40 minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event that k-wire, smooth, 1.4mmx100mm was alleged of issue s-11 (breakage during surgery) could not be confirmed, since the device returned for evaluation did not match the alleged failure mode.The product was returned for evaluation; however, during visual inspection, it was found to be in perfect condition.Both ends of the product was found to be in good condition.Considering the nature of the returned product and the alleged complaint, functional and dimensional inspection was not considered necessary.The device history record could not be reviewed as the lot number was not communicated.A review of the labeling did not indicate any abnormalities.If any further information is provided, the investigation report will be updated.
|
| |
|
Event or Problem Description
|
|
As reported: "k-wire was driven in to bone with stryker cd8 cordless driver and wire collet (facility owned).The k-wire broke off in the fibula.Case was an orif right fibula.The k-wire was barely protruding posterior and medial in the fibula.The capital fragment of the wire was removed but broken tip was stuck in the bone and unable to be removed easily.Tip protruding was too short to grip with the wire collet and further complicated by its position being difficult to access.Surgeon attempted to cut the tip to no avail.Surgeon attempted to bend the wire up and into the bone to prevent soft tissue irritation to no avail.Wire was ultimately removed by making multiple holes in the bone with a k-wire adjacent to the wire to loosen and free it to the point where it could be gripped and pulled free." delay in surgery and multiple holes made in the bone with another k-wire.Surgical delay of 30-40 minutes.
|
| |
|