| Model Number |
MICL13.7 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: product evaluation found that the lens was returned in liquid in a lens case/vial.Visual inspection found a piece of haptic missing.Work order search: no similar complaints types events were reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon attempted to implant a micl13.7,-10.0 diopter, implantable collamer lens in the patient's right eye (od) on (b)(6) 2017.The surgeon could not advance the lens in the injector.The surgeon believed that the lens could have been damaged and cause it to be stuck.There was no patient contact with the device.During the same surgery; a back-up lens was used and it was successful.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: date of birth corrected from (b)(6) to (b)(6).Type of reportable event corrected from serious injury to malfunction.Device evaluation: product evaluation found that the lens was returned in liquid in a lens case/vial.Visual inspection found the lens missing pieces of its haptic.(b)(4).
|
| |
|