• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR BONE ANCHORED SOLUTIONS AB BIA300 IMPLANT 3MM W ABUTMENT 6MM; COCHLEAR BAHA CONNECT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR BONE ANCHORED SOLUTIONS AB BIA300 IMPLANT 3MM W ABUTMENT 6MM; COCHLEAR BAHA CONNECT SYSTEM Back to Search Results
Model Number 92126
Device Problem Loss of Osseointegration (2408)
Patient Problem Pain (1994)
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
Per the clinic, the patient experienced pain and crusting at implant site.It was found the patient experienced a loss of osseointegration and was treated with a topical antibiotic for a duration of 2 weeks.It is unknown if there are plans to re-implant the patient with a new device as of the date of this report.
 
Manufacturer Narrative
The reported adverse is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Implant loss with baha implants can occur at any time following implantation.Known reasons for implant loss include lack of adequate bone quality/quantity, trauma, infection, generalised diseases and surgical complication.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIA300 IMPLANT 3MM W ABUTMENT 6MM
Type of Device
COCHLEAR BAHA CONNECT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
po box 82
mölnlycke, 435 2 2
SW  435 22
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
po box 82
mölnlycke, 43533
SW   43533
Manufacturer Contact
sujeewa wijesinghe
1 university avenue
macqaurie university, nsw 2109
AS   2109
MDR Report Key7069820
MDR Text Key93278607
Report Number6000034-2017-02225
Device Sequence Number1
Product Code LXB
UDI-Device Identifier09321502019514
UDI-Public(01)09321502019514(10)91160(17)160630
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100360
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number92126
Device Catalogue Number92126
Device Lot Number91160
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/23/2017
Initial Date FDA Received11/29/2017
Supplement Dates Manufacturer Received01/12/2018
Supplement Dates FDA Received02/06/2018
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-