ARTHREX INC. DISSECTOR, SJ 3.0MM X 7CM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
|
Back to Search Results |
|
Catalog Number AR-7300DS |
Device Problem
Metal Shedding Debris (1804)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/07/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was received and an evaluation was conducted.The complaint was confirmed / not confirmed.Device history record review revealed nothing relevant to this event.Complaint confirmed.The evaluation revealed metal gouging on the inner and outer tip areas of the shafts.Additionally, the cutting window of the outer shaft was damaged.Complainant's event is typically caused by excessive bending forces applied to the device during use.This is the first complaint of this type for this part/lot combination.The potential causes of this event are being communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
|
|
Event Description
|
It was reported that during an ankle arthroscopy case, metal shavings were being produced from the small joint dissector.It was indicated that most of the fragments were removed with the shaver/suction.A second small joint dissector was opened and used to complete the case.Patient is a male, (b)(6).Follow-up investigation: the metal shavings were noticed toward the end of the scope after, forward was used with the shaver.No x-ray was taken after the procedure to determine if there were any metal shavings remaining.Device is being returned for evaluation.
|
|
Search Alerts/Recalls
|
|
|