• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX INC. DISSECTOR, SJ 3.0MM X 7CM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX INC. DISSECTOR, SJ 3.0MM X 7CM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number AR-7300DS
Device Problem Metal Shedding Debris (1804)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/07/2017
Event Type  Injury  
Manufacturer Narrative
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was received and an evaluation was conducted.The complaint was confirmed / not confirmed.Device history record review revealed nothing relevant to this event.Complaint confirmed.The evaluation revealed metal gouging on the inner and outer tip areas of the shafts.Additionally, the cutting window of the outer shaft was damaged.Complainant's event is typically caused by excessive bending forces applied to the device during use.This is the first complaint of this type for this part/lot combination.The potential causes of this event are being communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
 
Event Description
It was reported that during an ankle arthroscopy case, metal shavings were being produced from the small joint dissector.It was indicated that most of the fragments were removed with the shaver/suction.A second small joint dissector was opened and used to complete the case.Patient is a male, (b)(6).Follow-up investigation: the metal shavings were noticed toward the end of the scope after, forward was used with the shaver.No x-ray was taken after the procedure to determine if there were any metal shavings remaining.Device is being returned for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISSECTOR, SJ 3.0MM X 7CM
Type of Device
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
ARTHREX INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath, adverse events
1370 creekside boulevard
naples, FL 34108-1945
8009337013
MDR Report Key7070102
MDR Text Key93293459
Report Number1220246-2017-00450
Device Sequence Number1
Product Code GFA
UDI-Device Identifier00888867041943
UDI-Public00888867041943
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 11/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/01/2020
Device Catalogue NumberAR-7300DS
Device Lot Number10005733
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/08/2017
Initial Date FDA Received11/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age17 YR
-
-