| Model Number |
3.3.2.0 |
| Medical Device Problem Codes |
Computer Software Problem (1112); Application Program Problem: Dose Calculation Error (1189); Data Problem (3196)
|
| Health Effect - Clinical Codes |
Hypoglycemia (1912); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/07/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A patch was created in version 3.3.4.4 which was released november 10, 2017.The effected clients will be upgraded as soon as possible.
|
| |
|
Event or Problem Description
|
|
After transition, transition basal dose may not be accounted for - patient transitioning from iv to subq treatment receives transitional basal dose via "give now" selection.At start of subq treatment, prompt may be provided to give full basal again.
|
| |
|