| Catalog Number |
9733881 |
| Medical Device Problem Codes |
Device Stops Intermittently (1599); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/31/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information not provided due to (b)(6) patient privacy regulations.No parts have been received by the manufacturer for evaluation.Part not returned for evaluation.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that during cranial resection procedure,the navigation system intermittently could not see the leds of the active head frame.The procedure was completed with the use of navigation.There was no delay to procedure.No impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
The cranial frame was returned to the manufacturer for analysis.Analysis found that when connected to a known good system, led #3 was not firing.Otherwise, the frame returned good geometry and divot error readings.Analysis found that the reported event was related to an electrical issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
|
| |
|