Model Number 425-97-007
Health Effect - Clinical Codes Failure of Implant (1924); No Code Available (3191)
Date of Event 11/10/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the stem being loose.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to the stem being loose.The in-vivo length of patient service for the implant was 22 days.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.This event is deemed to be non-product related.The root cause of this complaint was a revision surgery due to loosening.There are many factors that may contribute to the event that are outside the control of djo surgical are loose joints from inadequate soft tissue support, degenerative bone, patient bone deterioration, patient activities or trauma.Due to short time between previous and revision surgery, it is also possible that the event may have occurred due to improper surgical procedure or patient noncompliance with medical instructions.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPERFILL HIP
Common Device Name
TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 7
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key7070630
Report Number1644408-2017-01121
Device Sequence Number1186439
Product Code LPH
UDI-Device Identifier00888912078658
UDI-Public(01)00888912078658
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K130099
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/16/2023
Device Model Number425-97-007
Device Catalogue Number425-97-007
Device Lot Number172R1076
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/10/2017
Supplement Date Received by Manufacturer01/09/2018
Initial Report FDA Received Date11/30/2017
Supplement Report FDA Received Date01/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-32-004, LOT 892B1089
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
-
-