Model Number S7 |
Device Problems
Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/02/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.No parts have been returned to the manufacturer for evaluation.No parts have returned to the manufacturer.
|
|
Event Description
|
A site physician reported that, outside of a procedure, both screens flicker on the navigation system after it is powered on.There was no patient present when this issue was observed.
|
|
Manufacturer Narrative
|
A medtronic representative went to the site to test the equipment.The representative reported that the graphics card of the computer was cleaned to resolve the reported issue.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.
|
|