The reason for this revision surgery was due to a broken cam.The previous surgery and the revision detailed in this investigation occurred 7 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was a significant adverse event to the patient.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.The ultra-high molecular-weight-polyethylene (uhmw-pe) raw material was identified as ticona 1050, material lot 239541 and does not contain stearates.The poly was extruded and annealed to a 2.50"x3.75" rod and shown on the certificate of conformance to have a yield strength of 3234psi, an ultimate strength of 7700 psi, elongation to break of 421%, izod impact strength of 49.0 ft-lb/in and was certified to meet material specification #2 prior to shipment.As mentioned in a previous investigation: (test os9310-01, stability characteristics of the foundation posterior stabilized total knee design, pre-market evaluation, determined that the constraint of the design is primarily attributed to the post.It also determined that the design was similar to other designs showing the capacity to withstand medial/lateral shear forces in excess of normal knee reported maximums.Additionally, test 94-0001, fatigue testing of intercondylar post on foundation posterior stabilized tibial insert, determined that the intercondylar post of the foundation posterior stabilized tibial insert exhibited sufficient fatigue strength to withstand expected in-vivo loads.Wear mechanisms typically associated with total knee arthroplasty include abrasive wear, adhesive wear, third-body wear, delamination, and fatigue.High cyclic stresses are also known to produce delamination.Many factors can contribute to acceleration of wear mechanisms including patient factors, surgical technique, manufacturing methods, design and time in vivo.) customer complaint history showed two other complaints for this part number in which one involved a broken tibial post from another lot.There are no on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to a broken cam (post).The scope of this investigation is limited without having the parts available to djo surgical for evaluation.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
|