Catalog Number 367324
Device Problem Split (2537)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/05/2015
Event Type  malfunction  
Manufacturer Narrative
Bd received samples, representative and actual, from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for device breakage with the actual sample was observed.The indicated failure mode for the representative samples was not observed.An absolute root cause for this incident cannot be determined as bd was not able to duplicate the customer¿s indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.
 
Event Description
It was reported that a push button from a 23 g x.75 in.Bd vacutainer® winged safety push button blood collection set, split in two upon use.There was no report of injury or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® WINGED SAFETY PUSH BUTTON BLOOD COLLECTION SET
Type of Device
BLOOD COLLECTION SET
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7071873
MDR Text Key94203342
Report Number1710034-2017-00353
Device Sequence Number1
Product Code JKA
UDI-Device Identifier00382903673421
UDI-Public00382903673421
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K030573
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 11/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2017
Device Catalogue Number367324
Device Lot Number5177675
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/12/2015
Initial Date FDA Received11/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2017
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-