|
(b)(4).Batch # p9253k.Per photographic evaluation: upon visual inspection of the picture, two devices are present in the photograph, both devices appear to have been torn.Refer to physical analysis for details on investigation.Device analysis: the analysis results of the flr01 found that it was received with the retractor torn.The tear initiated 2 inches below the upper retractor ring and continued toward the lower retractor ring, causing the separation from the ring.No conclusion could be reached as to what may have caused the found condition.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.,a picture of the sample was received for analysis.The batch history record was reviewed and identified a scratches, nicks, or blemishes defect related to the reported incident were found.When this occurs, our quality system documents the necessary actions to ensure final product quality.In addition, no protocols or ncr related to the complaint, were found during the manufacturing process.The final quality release criteria were met before this batch was released for distribution.
|