The subject device was returned to olympus medical systems corp.(omsc) for evaluation.Omsc confirmed that the tip of the device was broken off.The manufacturing history was reviewed, with no irregularities noted.This type of the damage is most likely related to the operator's technique.Based on the similar cases in the past, it was known that the tip might be broken off since temperature of the cutting knife and tip increased while the high frequency output was activated.Also, it was known that temperature of the tip of the cutting knife increased because the electro surgical unit was used in the coagulation mode, the high-frequency output level was set too high, the activation time was too long, or because the contact length between the cutting knife and the tissue was too short.The instruction manual of the subject device warns; when the electrosurgical unit is used in the coagulation mode, deformation or break of the cutting knife and the tip could occur, for example when the high-frequency output is set too high or the length of the contact between the cutting knife and tissue is too short.During treatment, always ensure that the slider slides on the handle smoothly and that the electrosurgical knife observed in the endoscopic image is normal.
|
During a endoscopic submucosal dissection, the subject device was used.In the procedure, the distal part of the subject device was broken off and fallen into the patient.The fragment was not retrieved.The procedure was completed with another device.There was no patient injury reported.
|