Model Number KD-611L
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/07/2017
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus medical systems corp.(omsc) for evaluation.Omsc confirmed that the tip of the device was broken off.The manufacturing history was reviewed, with no irregularities noted.This type of the damage is most likely related to the operator's technique.Based on the similar cases in the past, it was known that the tip might be broken off since temperature of the cutting knife and tip increased while the high frequency output was activated.Also, it was known that temperature of the tip of the cutting knife increased because the electro surgical unit was used in the coagulation mode, the high-frequency output level was set too high, the activation time was too long, or because the contact length between the cutting knife and the tissue was too short.The instruction manual of the subject device warns; when the electrosurgical unit is used in the coagulation mode, deformation or break of the cutting knife and the tip could occur, for example when the high-frequency output is set too high or the length of the contact between the cutting knife and tissue is too short.During treatment, always ensure that the slider slides on the handle smoothly and that the electrosurgical knife observed in the endoscopic image is normal.
 
Event Description
During a endoscopic submucosal dissection, the subject device was used.In the procedure, the distal part of the subject device was broken off and fallen into the patient.The fragment was not retrieved.The procedure was completed with another device.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE ELECTROSURGICAL KNIFE
Type of Device
SINGLE USE ELECTROSURGICAL KNIFE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
katsuaki morita
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key7073300
MDR Text Key94649619
Report Number8010047-2017-01881
Device Sequence Number1
Product Code KNS
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK092309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Other
Type of Report Initial
Report Date 12/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberKD-611L
Device Lot NumberK6916
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2017
Date Manufacturer Received11/08/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-