• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. MERLIN PROGRAMMER; PACEMAKER PROGRAMMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. MERLIN PROGRAMMER; PACEMAKER PROGRAMMER Back to Search Results
Model Number 3650
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems Electric Shock (2554); No Known Impact Or Consequence To Patient (2692)
Event Date 11/10/2017
Event Type  malfunction  
Event Description
During device preparation, while pushing the start button, the programmer gave the user an electric shock and then would not power on.This happened twice to different users.The user's were stable with no adverse consequences.Related manufacturer report number: 2017865-2017-35232.
 
Manufacturer Narrative
The programmer was returned for analysis.Visual inspection revealed no physical damage to the unit.During analysis, the programmer was unable to advance past the startup page.The root cause of the start-up anomaly was identified as a defective power switch.However, no root cause was determined for the electric shock to the user in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
PACEMAKER PROGRAMMER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7076710
MDR Text Key93634141
Report Number2017865-2017-35302
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Other Device ID Number05414734500661
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/03/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-