• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ETHICON; HARMONIC SCALPEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ETHICON; HARMONIC SCALPEL Back to Search Results
Model Number HARHD36 1000ISHEARS
Device Problems Melted (1385); Device Displays Incorrect Message (2591)
Patient Problem Injury (2348)
Event Date 11/16/2017
Event Type  Injury  
Event Description
During laparoscopic hiatal hernia repair using harmonic scalpel, the scalpel malfunctioned.The machine alerted the surgical team, scalpel was withdrawn and examined, the plastic jaws of the grasper appeared to have been melted.Instrument removed from the field and a new one obtained.To evaluate the esophagus and stomach a gastroscope was used, while blanching of the anterior left lateral segment of the distal esophagus was noted (concern for cautery injury.) scope then passed to the stomach and a gastrotomy injury was identified as well.Scope with drawn, distal esophageal lesion reevaluated, bubbling seen from the laparoscopic view and a tear was identified.Converted to an open procedure and exploratory lap done with repair of esophageal tear and gastrotomy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHICON
Type of Device
HARMONIC SCALPEL
Manufacturer (Section D)
ETHICON
MDR Report Key7078795
MDR Text Key93735868
Report NumberMW5073695
Device Sequence Number1
Product Code NLQ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2022
Device Model NumberHARHD36 1000ISHEARS
Device Lot NumberP93KOU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age55 YR
Patient Weight72
-
-