|
Clinical evaluation performed by medical affairs director on 24 november 2017.Six months after primary surgery, a screw of a cervical plate broke in the vertebral body.The cause for fracture is probably delayed fusion; all these implants are meant to carry load only temporarily.As the implant has not been retrieved, we could not analyze it in order to verify if a manufacturing defect is present.The possibility of fusion for the patient is probably hindered but not definitely compromised; at 9 months, it seems improved but clinical analysis is not available.Batch review performed on 01 december 2017: lot 141344: 104 items manufactured and released on 22-oct-2014.Expiration date: 30- sept- 2019.No anomalies found related to the problem.To date, 32 items of the same lot have been already sold without any similar reported event.
|