Catalog Number CDC-44052-VPS2 |
Device Problems
Failure to Advance (2524); Split (2537)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/17/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
The customer alleges that the sheath flowered making it unable to advance into the patient.Clinician completed procedure by way of seldinger technique.
|
|
Manufacturer Narrative
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
|
Event Description
|
The customer alleges that the sheath flowered making it unable to advance into the patient.Clinician completed procedure by way of seldinger technique.
|
|