• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. ALTIVATE SHOULDER; ALTIVATE ANATOMIC, SHORT HUMERAL STEM, 14MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. ALTIVATE SHOULDER; ALTIVATE ANATOMIC, SHORT HUMERAL STEM, 14MM Back to Search Results
Model Number 520-14-000
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/09/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient being presented with a torn subscap repair status post operation of a total shoulder.The surgeon decided to go in and replace all the implants from the total shoulder and do a reverse shoulder.
 
Additional Manufacturer Narrative
The reason for this revision surgery was a torn subscap.The in-vivo length of patient service for the implant was 6 months.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.This event is deemed to be non-product related.The root cause of this complaint was a revision surgery due to the subscap tear.There are multiple factors that may contribute to the event that are outside the control of djo surgical are soft tissue impingement, excessive range-of-motion, excessive load, patient non-compliance with medical instructions, patient activities or trauma.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTIVATE SHOULDER
Common Device Name
ALTIVATE ANATOMIC, SHORT HUMERAL STEM, 14MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key7080377
Report Number1644408-2017-01092
Device Sequence Number1260253
Product Code KWS
UDI-Device Identifier00190446176039
UDI-Public(01)00190446176039
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K162024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/03/2023
Device Model Number520-14-000
Device Catalogue Number520-14-000
Device Lot Number853U1007
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/09/2017
Supplement Date Received by Manufacturer12/22/2017
Initial Report FDA Received Date12/04/2017
Supplement Report FDA Received Date01/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-07-000, LOT 888U1030; 520-50-218, LOT 949U1007; 521-07-250, LOT 893U1008
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
-
-