MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP SHOULDER; SOCKET, INSERT 40MM STD. RSP HUMERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP SHOULDER; SOCKET, INSERT 40MM STD. RSP HUMERAL Back to Search Results
Model Number 508-00-040
Health Effect - Clinical Code Pain (1994)
Date of Event 11/17/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient complaining of posterior pain.The surgeon found that the patient's x-rays, computerized axial tomography (cat) scans and magnetic resonance imaging (mri) looked good.The surgeon did the revision and indicated the implants looked fine too.He did not know why the patient was feeling pain and didn't think anything would change after the revision surgery.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the patient complained of posterior pain.The in-vivo length of patient service for the implant was 11 months.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.This event is deemed to be non-product related.The root cause of this complaint was a revision surgery due to the pain.There are many factors that may contribute to the event that are outside the control of djo surgical are inadequate soft tissue support, degenerative bone, excessive loading, patient non-compliance with medical instructions or patient activities.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Common Device Name
SOCKET, INSERT 40MM STD. RSP HUMERAL
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key7087167
Report Number1644408-2017-01103
Device Sequence Number1466513
Product Code KWS
UDI-Device Identifier00888912144193
UDI-Public(01)00888912144193
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K051075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Model Number508-00-040
Device Catalogue Number508-00-040
Device Lot Number857C1900
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/17/2017
Supplement Date Received by Manufacturer01/01/2018
Initial Report FDA Received Date12/06/2017
Supplement Report FDA Received Date01/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
508-40-101. LOT 875C1113
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
-
-