| Model Number |
N/A |
| Medical Device Problem Codes |
Detachment Of Device Component (1104); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
10/31/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The international customer reported that the cook-swartz doppler probe crystal came off prematurely at the end of surgery.The operator had to replace the crystal and re-open the patient's neck.The circumstances surrounding the usage and handling of the device leading up to the product malfunction were not reported.The device is reportedly available for return; however, as of the date of this report, no device has yet been received for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation/evaluation summary: the device was inspected under magnification and confirmed the customer's complaint that the cuff became detached from the crystal.A functional test was performed on the probe and an audible pulse sound was verified.No other nonconformities or anomalies were observed on the device.A review of the device history record and manufacturing and quality control records was conducted.There is no evidence that defective product was manufactured using current drawings and device specifications.Complaints will continue to be monitored according to established complaint monitoring procedures.
|
| |
|