| Catalog Number |
71442109 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/20/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that even though the orthomatch universal primary cutting block was attached firmly the alignment was incorrect resulting in a valgus determination on the femoral cuts and poor inclination on the tibial cuts.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint device was not returned for evaluation please review attached for investigation results.
|
| |
|
Additional Manufacturer Narrative
|
|
The universal cut block was returned and evaluated.A visual inspection found that the device had sharp burs around both sides of the cutting guide.Some abrasions were also noticed throughout the device.Additionally, a dimensional inspection was attempted; however the damage and deformation at several features of the device would not allow for accurate measurement.Finally, clinical investigation concluded the following: without sufficient clinically relevant supporting documentation, a thorough medical assessment and/or causal relationship cannot be provided.The patient impact beyond possible instability from the reported poorly maintained flexion/extension gap and the 45 minute surgical extension cannot be concluded.No further medical assessment is warranted at this time.We recommend that all reusable instruments be routinely inspected for wear and damage so they can be replaced as necessary.Reviewing the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Based on this investigation, the need for corrective action is not indicated.No additional actions are being taken at this time; however we will continue to monitor for future complaints and investigate further as necessary.
|
| |
|
Search Alerts/Recalls
|