• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC RADIAL ARTERY PRESSURE MONITORING SET; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC RADIAL ARTERY PRESSURE MONITORING SET; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number N/A
Medical Device Problem Codes Difficult to Flush (1251); Unable to Obtain Readings (1516)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/27/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.
 
Event or Problem Description
The international customer reported that, when the radial artery pressure monitoring set was employed, the waveform dampened immediately following insertion.Flushing and wrist repositioning improved the signal, but 48 hours post insertion, the waveform was ultimately completely absent, and flushing was difficult.The blood draw had become slow, and significant pressure was required at the insertion site.The waveform was again absent at the 54 hour post insertion.Flushing and blood draw became impossible.Additional information has been requested from the customer, but has not yet been provided.The device is reportedly available for return and investigation; however, as of the date of this report, no device has yet been received for evaluation.
 
Additional Manufacturer Narrative
Previously requested additional information was received from the customer.What types of fluids were being infused through the line? heparinized saline.What is the maintenance protocol for the catheter? heparinized saline under pressure to flow at 3mls/hr.Can patient demographics be provided? no ¿ reason is foip.Please describe the other products and equipment used during the procedure (endoscope type and model, introducer, wire guide, etc.) standard stylet that comes with the catheter to assist with insertion.Did the patient have a pre-existing condition and/or pre-diagnosis relevant to this event? no.Please describe the placement site for this procedure.As described in the description of problem it was placed in the radial artline in the wrist did any section of the device remain inside the patient¿s body? no ¿ question is not relevant to this problems description.Did the patient require any additional procedures due to this occurrence? no ¿ question is not relevant to this problem did the event result in a death? no.Investigation ¿ evaluation: a review of the complaint history, device history record, documentation, and quality control data of the device was conducted during the investigation.No complaint device has been returned for investigation.No photos have been provided.However, a documentation investigation/evaluation was conducted.A review of the device history record found no other non-conformances associated with the complaint device lot number; 7970203.Additionally, a review of the manufacturer's complaint database found no other complaints associated with the complaint device lot number; 7970203.Based on the investigation evaluation, there is no indication that a design or process related failure mode contributed to this event.Current controls for manufacturing are in place to assure functionality and device integrity prior to shipping.Review of production and quality documentation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.A definitive root cause could not be determined, however; based on the provided information the likely root cause is deemed to be; ¿inconclusive¿.Per the quality engineering risk assessment no further action is required.We will continue to monitor similar complaints and have notified the appropriate personnel of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIAL ARTERY PRESSURE MONITORING SET
Common Device Name
DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7088670
Report Number1820334-2017-04164
Device Sequence Number1466542
Product Code DQO
UDI-Device Identifier00827002019089
UDI-Public(01)00827002019089(17)200606(10)7970203
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberC-PMS-300-RA
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 11/14/2017
Supplement Date Received by Manufacturer02/15/2018
Initial Report FDA Received Date12/06/2017
Supplement Report FDA Received Date03/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-