Model Number 520-01-242
Health Effect - Clinical Codes Failure of Implant (1924); No Code Available (3191)
Date of Event 11/20/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the glenoid components coming loose over time.The patient had a total shoulder arthroplasty (tsa) procedure in 2012.The surgeon removed the original components and converted to a reverse shoulder prosthesis (rsp).
 
Additional Manufacturer Narrative
The reason for this revision surgery was the glenoid components came loose.The in-vivo length of patient service for the implant was 5 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component(s) used in the previous surgery met design and manufacturing requirements.There were no ncmrs associated with the product(s) that may have contributed to the reported event.The device and its applicable concomitant devices were verified to have gone through acceptable sterilization processes and were within their respective expiration dates at the time of use during the previous surgery.Customer complaint history of the reported device(s) showed no present trends or on-going issues that are in need of review.This event is deemed to be non-product related.The root cause of this complaint was a revision surgery due to the components becoming loose.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
TURON PEGGED GLENOID IMPLANT, SIZE 42
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key7088777
Report Number1644408-2017-01113
Device Sequence Number16510545
Product Code KWS
UDI-Device Identifier00888912024952
UDI-Public(01)00888912024952
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K123982
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/28/2017
Device Model Number520-01-242
Device Catalogue Number520-01-242
Device Lot Number191G1033
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/20/2017
Supplement Date Received by Manufacturer01/02/2018
Initial Report FDA Received Date12/06/2017
Supplement Report FDA Received Date01/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-000, LOT 878C1123; 520-01-010, LOT 455G1051; 520-42-016, LOT 909C1017
Outcome Attributed to Adverse Event Required Intervention;
Patient Age84 YR
-
-