MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPTA QUAD CRTP; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION Back to Search Results
Model Number W4TR01
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/23/2017
Event Type  Injury  
Event Description
It was reported that a full power on reset was observed during a routine device follow up.The patient was not around any sources of high energy at the time.The device was reset and remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated a software error.Device circuit error occurred on (b)(6)2017.Analysis of the device memory indicated a full electrical reset.Analysis of the device memory had an observation relating to the battery longevity estimator.Longevity unavailable, stating "remaining longevity is re-initializing.Values will be available within 1 week." if information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEPTA QUAD CRTP
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7092744
MDR Text Key94065383
Report Number3004209178-2017-25306
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00643169735644
UDI-Public00643169735644
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/14/2019
Device Model NumberW4TR01
Device Catalogue NumberW4TR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
459888 LEAD, 5076-58 LEAD, 5076-52 LEAD
Patient Outcome(s) Required Intervention;
Patient Age57 YR
-
-