| Model Number |
52516-15 |
| Medical Device Problem Code |
Calibration Error (1078)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
| Date of Event |
10/27/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Lot review: no exception documents cited in work order.
|
| |
|
Event or Problem Description
|
|
Optical module of catheter was not reading svo2.The cco box was changed out but same issues occurred.Facility reports that there was a critical delay in therapy and medical intervention was required.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Lot review: no exception documents cited in work order.Visual analysis: received one used 52516-15.There is residual blood observed but was flushed out of the catheter during decontamination.The proximal tubing is occluded.No visual anomalies.Functional analysis: initial calibration of the returned device was successful.It was observed that crosstalk between the two optic fibers was occurring.Crosstalk could possibly lead to inaccurate svo2 readings.Summary conclusion: the report of svo2 not reading could not be replicated.No defect was found and the heating coil was working within specifications.
|
| |
|
Event or Problem Description
|
|
Optical module of catheter was not reading svo2.The cco box was changed out but same issues occurred.Facility reports that there was a critical delay in therapy and medical intervention was required.No other information was provided.
|
| |
|