• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. PS KNEE; INSERT, POSTERIOR STABILIZED, SERIES 500, SIZE8 11MM CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. PS KNEE; INSERT, POSTERIOR STABILIZED, SERIES 500, SIZE8 11MM CM Back to Search Results
Model Number 360-11-508
Health Effect - Clinical Code Discomfort (2330)
Date of Event 11/20/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient having anterior knee discomfort.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to the patient's discomfort.The previous surgery and the revision detailed in this investigation occurred 5 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There was a non-conforming material report (ncmr) # 13842 associated with the part 360-11-508, insert, posterior stabilized series 500 size8 11mm.The non-conforming material report documented a nonconformance that out of 10 quantity lot 1 item was rejected due to multiple scratches greater than 1 micron on condyle's and was scraped.Also, all other items in the lot met the fit, design and functional requirements.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to patient's discomfort.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are multiple factors that may contribute to the event that are outside the control of djo surgical such as: loose joint from inadequate soft tissue, excessive range-of-motion, weight of the patient, degenerative bone disease, scar tissue, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS KNEE
Common Device Name
INSERT, POSTERIOR STABILIZED, SERIES 500, SIZE8 11MM CM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7095748
Report Number1644408-2017-01128
Device Sequence Number1873848
Product Code JWH
UDI-Device Identifier00888912116350
UDI-Public(01)00888912116350
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K933539
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2015
Device Model Number360-11-508
Device Catalogue Number360-11-508
Device Lot Number59600418
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/20/2017
Supplement Date Received by Manufacturer01/10/2018
Initial Report FDA Received Date12/07/2017
Supplement Report FDA Received Date01/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
-
-