• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC STARDRIVE SCREWDRIVER SHAFT T4/42MM/SELF-RETAINING; SCREWDRIVERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC STARDRIVE SCREWDRIVER SHAFT T4/42MM/SELF-RETAINING; SCREWDRIVERS Back to Search Results
Catalog Number 03.114.002
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/07/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient weight was not provided for reporting.Device is an instrument and is not implanted / explanted.Device is not expected to be returned for manufacturer review/investigation.Device evaluation/investigation could not be completed; no conclusion could be drawn as product was not returned to manufacturer.Device history records review could not be completed without lot number.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was report that on (b)(6) 2017, patient underwent implant surgery for treatment of a (orif) open reduction internal fixation to the right fourth metacarpal bone due to a bone fracture where the surgeon was using the 1.5 mm (lcp)locking compression system.During the produce, surgeon chose to use one (1) 1.5 mm lcp t-plate and three (3) 1.5 mm cortex self tapping screws.Surgeon originally wanted to use a condylar plate for the fracture reduction, but this plate was unavailable in the lcp set.(due to this plate was used during a surgery the day before.) surgeon cut and contoured the t-plate to the appropriate length and attempted to insert three (3) cortex screws.Surgeon struggled with three (3) different screwdrivers when attempting to implant each of the three (3) cortex screws.All three screwdrivers became stripped and would no longer retain the screws.Additional screwdrivers were available in the operating room to continue the surgery.At this point in the procedure, surgeon removed the previously implanted three cortex screws and plate.Patient was ultimately implanted with two (2) 1.5 mm cortex screws and fiber wire sutures to compress the bone fracture.Surgery was completed successfully with a 45-minute time delay.Patient was reported in stable condition.Concomitant devices reported: 1.5 mm cortex screws slf-tpng with t4 stardrive recess 10 mm (part# 02.214.110, lot# unknown, quantity 2) 1.5 mm cortex screws slf-tpng with t4 stardrive recess 12 mm (part# 02.214.112, lot# unknown, quantity 1).This report is for one (1) stardrive screwdriver shaft t4/42 mm/self-retaining.This is report 2 of 4 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STARDRIVE SCREWDRIVER SHAFT T4/42MM/SELF-RETAINING
Common Device Name
SCREWDRIVERS
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7096174
Report Number1719045-2017-50013
Device Sequence Number16369608
Product Code HXX
UDI-Device Identifier10886982072993
UDI-Public(01)10886982072993(10)LOT#UNKNOWN
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 11/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.114.002
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/07/2017
Initial Report FDA Received Date12/07/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
-
-