INTUITIVE SURGICAL,INC. DA VINCI XI SURGICAL SYSTEM; ENDOSCOPIC INSTRUMENT CONTROL SYSTEM Back to Search Results
Model Number IS4000 A70P7B
Device Problems Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/16/2017
Event Type  malfunction  
Manufacturer Narrative
The isi fse replaced two fiber cables to resolve the issue experienced by the customer.However, the fiber cables will not be returned to isi for failure analysis as these cables were scrapped in the field.This complaint is being reported due to a da vinci system malfunction rendering the da vinci system unavailable for use after the start of a surgical procedure.Although no patient harm occurred, if this malfunction were to recur it could cause or contribute to an adverse event.
 
Event Description
It was reported that during a da vinci-assisted surgical procedure, the system generated a continuous non-recoverable 319 fault.The customer attempted troubleshooting by removing instruments and performing two system reboots with no change.The customer stated that one of the instruments had to be removed using an emergency release wrench.The intuitive surgical, inc.(isi) technical support engineer (tse) was able to confirm the 319 errors in the logs reported by the patient cart gantry indicating nodes not responding.The tse continued to troubleshoot with the customer but the issue persisted.At that time, the customer made the decision to convert and complete the procedure using traditional laparoscopic techniques.There was no report of patient harm, adverse outcome, or injury.An isi field service engineer (fse) was dispatched to the facility and was able to reproduce the reported failure.The fse replaced the two fiber cables to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DA VINCI XI SURGICAL SYSTEM
Type of Device
ENDOSCOPIC INSTRUMENT CONTROL SYSTEM
Manufacturer (Section D)
INTUITIVE SURGICAL,INC.
sunnyvale CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC.,
950 kifer rd
sunnyvale CA
Manufacturer Contact
tabitha reed
950 kifer rd
sunnyvale, CA 
4085232420
MDR Report Key7096459
MDR Text Key94947585
Report Number2955842-2017-00820
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 11/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIS4000 A70P7B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/16/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-